Unit Weight |
63.0 kg (140.0 lb) |
Method |
ASTM D445/D446, IP 71, ISO 3104/3105, SAE J300 |
Dimensions (W x D x H) |
36.0 cm × 66.0 cm × 72.0 cm (14.3 in × 26.0 in × 28.5 in) |
Shipping Dimensions (W x D x H) |
81.3 cm × 53.3 cm × 99.0 cm (32.0 in × 21.0 in × 39.0 in) |
Shipping Weight (With All Items) |
96.6 kg (213.0 lb) |
Maximum Throughput |
24 tests per hour |
Sample Capacity |
28 positions (2 x 14 positions), automated |
Viscosity Range |
0.5 mm²/s (cSt) to 10,000 mm²/s (cSt) in 100-fold increments (depending on viscometer tube selection). Fast run tubes are also available. |
Timing Resolution |
0.01 s (timing accuracy to ± 0.001 s) |
Temperature Range and Accuracy |
20 °C to 100 °C,± 0.01 °C (temperatures within 5 °C of ambient and below require bath cooling)
Up to 150 °C,± 0.03 °C (with high temperature bath)
Down to 15 °C,± 0.03 °C (with TE cooling)
|
Minimum Sample/Solvent Volume |
8 mL sample/15 mL solvent per test (as little as 3 mL with fast run tubes) |
Operating Conditions |
15 °C to 30 °C, 10% to 75% relative humidity (non-condensing),Installation Category II; Pollution Degree 2 |
Electrical Specifications |
100 Vac, 50/60 Hz; 115 Vac, 50/60 Hz; 230 Vac, 50/60 Hz; 1200 watts power consumption |
Compliance |
CE Mark; EMC directive (2004/108/EC); Low voltage directive (2006/95/EC); HI-POT (1900 Vdc,60 sec.); ROHS |
Additional Standard Features |
Internal heated waste drain lines,dual solvent input |
Data Output |
USB |